ADSC Quantum 210 CCD detector Specification |
||||||
- Protein crystallographyMain steps:- Protein purification- Crystallisation Special:- Programs for crystallography- X-ray detectors - ADSC - Quantum 1 - Quantum 4 - Quantum 210 - Quantum 315 Basic tutorials:- Chemistry- Protein - Peptide - Amino Acids Xtal community:- CCP4BB |
Quantum 210 CCD X-ray detector general description
Quantum 210 CCD X-ray detector is a two by two array CCD detector, consisting of four modules. The total active area is approximately 44100 mm2 (210 x 210 mm) with no more than 400 m slit between detecting areas.Total number of pixels = 4096 x 4096 with final image size about 16.8 Mb. CCD chip specification
Each individual Thomson THX7899 (2K x 2K) chip with 2048 x 2048 pixels (all of them are used) is connected via fiber optic taper ratio is 3.6 to 1 to 105 x 105 mm X-ray sensitive phosphor area, optimized for 1 Å radiation. This mean that each pixel on CCD chip corresponds to 51µm square at the X-ray sensitive area. ADSC Quantum 210 Detector System Overview
Standard installation package include:
ADSC Software for Quantum 210 CCD detectorQuantum 210 CCD detector is supplied with ADSC software package. ADSC software has modular design, which allowing use independent process via TCP/IP network with central Graphics User Interface (GUI) control. ADSC software with Graphics User Interface allow to make simple, safe control of data collection with estimation of the full data collection time and display collected images.ADSC Warranty:ADSC warrants both parts and labor for any defects due to faulty material or workmanship for a period of two years from date of installation.For more details please visit official web-site: http://www.adsc-xray.com/ |
|||||